Synopsis
Explain how a sealed negative-pressure system manages exudate and wound geometry, select a suitable wound, protect vulnerable structures and judge benefit from measurable progress rather than device presence.
- NPWT places porous interface material in a prepared wound, seals it with an adhesive drape and applies controlled sub-atmospheric pressure through tubing to a canister.
- It removes exudate, reduces oedema in selected tissue, draws wound surfaces together and supports granulation, but cannot replace debridement, perfusion or source control.
- For the cited V.A.C. device, never place foam directly over exposed blood vessels, anastomoses, organs or nerves; protect vulnerable structures with the indicated barrier and specialist technique before applying negative pressure.
- Count and document every piece of foam at insertion and reconcile the count at removal. Before selecting this device, address its contraindications: malignancy in the wound, untreated osteomyelitis, unexplored or non-enteric fistula and necrotic tissue with eschar require another or specialist plan.
- NICE does not recommend routine NPWT for pressure ulcers except when needed to reduce dressing changes, such as with heavy exudate.
- Judge benefit by wound dimensions, granulation, exudate, seal integrity, pain and the next closure goal; stop when it no longer advances that goal or harm outweighs benefit.
Key red flags
Active uncontrolled bleeding, exposed unprotected vessel or organ, or sudden bright blood in the canister requires stopping therapy and immediate surgical assessment.
A retained foam fragment, blocked tubing, lost seal or pump-off interval defeats therapy and can promote infection or tissue injury.
Necrotic eschar, untreated deep infection or an enteric fistula without a specialist plan requires source assessment before routine NPWT.